Welcome to BeaconMedaes
Home >> NFPA Products >> Vacuum >> Claw >> O2 Assured with VSD

Claw Vacuum Systems – O2 Assured with VSD

The Most Efficient Vacuum Technology Available

The BeaconMedæs Claw Vacuum System utilizes a dry, frictionless, multi-claw technology. The rotors (claw) spin in opposite directions, synchronized via precision gears, creating a highly efficient vacuum system. The rotors never contact the housing or each other, which removes the need for any lubricant or sealing fluid inside the pumping chamber.

Variable Speed Drive

Variable speed drive matches the vacuum motor speed to the vacuum demand, resulting in maximum efficiency of the vacuum system.

  • Greater energy efficiency
  • Vacuum level consistency
  • Reduced noise and heat
  • Longer machine life

Low Maintenance

Due to the contact-less operation, the only maintenance is a simple matter of replacing the gear lube and changing an inlet filter annually.

Safe Choice for Waste Anesthesia Service

The Claw is uniquely available for oxygen and WAGD service.  We label it “O2 Assured” and guarantee that it will safely operate under high oxygen concentration.


Click here to download product documents for O2 Vertical VSD 60 Hz
Click here to download product documents for O2 SPC VSD 60 Hz
Click here to download product documents for O2 Modular VSD 60 Hz
Click here to download product documents for O2 Vertical VSD 50 Hz
Click here to download product documents for O2 SPC VSD 50 Hz
Click here to download product documents for O2 Modular VSD 50 Hz

Model HP KW System System Capacity1
SCFM l/s
60 Hz 50 Hz 60 Hz 50 Hz
Vertical SPC (Single Point Connection) Base Mount Systems with VSD
VAV05D-060H-D_V 5.4 4 Duplex 38 32 17.9 15.1
SPC (Single Point Connection) Base Mount Systems with VSD
VAS05D-120V-D_V 5.4 4 Duplex 38 32 17.9 15.1
VAS06D-120V-D_V 6.4 4.8 52 43 24.5 20.3
VAS07D-120V-D_V 7.4 5.5 65 54 30.7 25.5
VAS08D-120V-D_V 8.7 6.5 77 64 36.3 30.2
VAS05T-120V-T_V 5.4 4 Triplex 76 64 35.9 30.2
VAS06T-200V-T_V 6.4 4.8 104 86 49.1 40.6
VAS07T-200V-T_V 7.4 5.5 130 108 61.4 51
VAS08T-200V-T_V 8.7 6.5 154 128 72.7 60.4
VAS06Q-200V-Q_V 6.4 4.8 Quadruplex 156 129 73.6 60.9
VAS07Q-200V-Q_V 7.4 5.5 195 162 92.0 76.4
VAS08Q-200V-Q_V 8.7 6.5 231 192 109.0 90.6
Expandable SPC (Single Point Connection) Base Mount Systems with VSD
VAS05D-120V-T_V 5.4 4 Duplex Expandable to Triplex 38 32 17.9 15.1
VAS06D-200V-T_V 6.4 4.8 52 43 24.5 20.3
VAS07D-200V-T_V 7.4 5.5 65 54 30.7 25.5
VAS08D-200V-T_V 8.7 6.5 77 64 36.3 30.2
VAS06T-200V-Q_V 6.4 4.8 Triplex Expandable to Quadruplex 104 86 49.1 40.6
VAS07T-200V-Q_V 7.4 5.5 130 108 61.4 51
VAS08T-200V-Q_V 8.7 6.5 154 128 72.7 60.4
Modular Base Mount Systems with VSD
VAM05D-120V-D_V 5.4 4 Duplex 38 32 17.9 15.1
VAM06D-120V-D_V 6.4 4.8 52 43 24.5 20.3
VAM07D-120V-D_V 7.4 5.5 65 54 30.7 25.5
VAM08D-120V-D_V 8.7 6.5 77 64 36.3 30.2
VAM05T-120V-T_V 5.4 4 Triplex 76 64 35.9 30.2
VAM06T-200V-T_V 6.4 4.8 104 86 49.1 40.6
VAM07T-200V-T_V 7.4 5.5 130 108 61.4 51
VAM08T-200V-T_V 8.7 6.5 154 128 72.7 60.4
VAM05Q-200V-Q_V 5.4 4 Quadruplex 114 86 53.8 40.6
VAM06Q-200V-Q_V 6.4 4.8 156 129 73.6 60.9
VAM07Q-200V-Q_V 7.4 5.5 195 162 92.0 76.4
VAM08Q-200V-Q_V 8.7 6.5 231 192 109.0 90.6
Expandable Modular Base Mount Systems with VSD
VAM05D-120V-T_V 5.4 4 Duplex Expandable to Triplex 38 32 17.9 15.1
VAM06D-200V-T_V 6.4 4.8 52 43 24.5 20.3
VAM07D-200V-T_V 7.4 5.5 65 54 30.7 25.5
VAM08D-200V-T_V 8.7 6.5 77 64 36.3 30.2
VAM05T-200V-Q_V 5.4 4 Triplex Expandable to Quadruplex 76 64 35.9 30.2
VAM06T-200V-Q_V 6.4 4.8 104 86 49.1 40.6
VAM07T-200V-Q_V 7.4 5.5 130 108 61.4 51
VAM08T-200V-Q_V 8.7 6.5 154 128 72.7 60.4
1. All capacities are shown as NFPA system capacities (reserve vacuum pump on standby). All capacities are measured at 19" Hg.
2. Normal operating conditions at a maximum ambient of 105° F. Consult factory for higher ambient conditions.


Click Here To View Large Image
Claw2.jpg

Global Directory

Global Directory

Global Directory

Global Directory